DXC100: Why are the temperature readings on my DXC100 higher than expected and the humidity readings lower than expected?
The DXC100 has been replaced by the DXC120. This answer is for owners of a DXC100.
Humidity and temperature measurements are internal measurements, i.e. they are taken inside the case. Therefore, the temperature is higher inside the case than outside, and conversely, the humidity is lower inside the case than outside. These measurements must be accurate and taken inside the case, as they are used in the automated calibration process of the product.
To apply an offset to these measurements, one option is to use the mathematical channels available in DracalView to create a new channel to which you apply your 1-point compensation (i.e. an offset).